Page 55 - Home Care Catalogue Edition 4A
P. 55

Choosing a nebuliser
HOT TopTM2 Tracheobronchial deposition (2–5μm) high speed nebulisation for greater patient compliance
CirrusTM2 Tracheobronchial deposition (2–5μm)
Particles smaller than 0.5 microns expired back to atmosphere
Oral cavity
Particles above 5–10 microns impact on the upper respiratory tract
Nasal cavity
Pharynx Larynx Trachea
10 8 86
10 64
2
Tracheobronchial deposition
42
10 10 88
66 44
22
HOT Top2 nebuliser
The HOT Top2 is designed for high speed nebulisation. It signi cantly reduces treatment time compared with conventional nebulisers.
At a driving gas  ow of 8 L/min, 80% of the volume output will be particles of less than 5 microns in diameter with a Mass Median Diameter (MMD) of 3 microns.
Cirrus2 nebuliser
Neonatal Therapy Training Video available at www.intersurgical.com
Nebuliser Therapy • Nebulisers 55
The Cirrus2 has been designed to minimise the residual volume and reduce drug wastage.
At a driving gas  ow of 8 L/min, 74% of the volume output will be particles less than 5 microns in diameter with a MMD
of 3.3 microns.